FREESTYLE XANA PFAS

Based on the robust and established FREESTYLE XANA robotic system for large-volume water samples (0.1 L – 10 L), LCTech has developed a solution specifically for PFAS applications, in which plastics containing fluorine such as PTFE are avoided.

Further PFAS Robotic Systems:

    DETAILS AT A GLANCE

    XANA for Automated SPE in PFAS Analysis

    09-FREESTYLE-XANA-PFAS@2x sm
    1 SPE Module
    2 Processing Station
    3 External Processing Station
    4 Sample Rack
    5 EVAporation

    The column is firmly connected to the arm, therefore sample and solvent can be dosed precisely and at a constant flow rate via the SPE column with positive pressure using a syringe pump. Drying is also carried out via positive pressure with nitrogen.

    • Non-stop Policy: If a column becomes blocked, the system detects the overpressure. The column is reset, the system cleans itself automatically and continues with the processing of the next column
    • One rack for all types of columns: reusable adapters allow easy changing and mixing of columns of any standard format
    • Columns can be moved to any position in the rack area
    • Three solvents as standard, upgrade option up to six or even fifteen solvents

    In the processing station on the FREESTYLE platform, 3 SPE columns are processed simultaneously in 2 blocks: In block 1, 3 columns are conditioned, loaded and washed in parallel, while in the following block 2, 3 columns are dried in parallel.

    The following videos show the processing station of the FREESTYLE XANA. They are identical for all XANA models:

    The external processing station is responsible for the fluidics management of the water module with the pumping mechanics of up to 10 L samples, the valves and the sensor technology for process monitoring.

    All components in the fluidic way ( tubes, valves, ...) are PTFE-free!

    The sample rack offers space for 24 x 1 L sample bottles; optionally 3 sample containers with up to 10 L volume each can be connected to the system. 

    Automatic rinsing of the sample bottles possible for optimal quantitative transfer of the sample.

    The vacuum concentrator D-EVA allows unattended, parallel and fast concentration of several samples with automatic stop at a defined volume between 10 and 150 µL.

    Read more

    Specifications

    Max. Volume per sample

    up to 4 L (10 L retrofitting possible)


    Samples per sequence

    24


    Duration per sequence

    approx. 8 hours (for US EPA 537.1)


    Samples simultaneously

    up to 6


    Location

    on bench supplied


    Fume hood

    not required


    PFAS compatibility

    PE tubings, PTFE-free syringe, PTFE-free valves


    Solvent compatibility

    all necessary solvents according to the regulations are compatible


    Assignment of LCTech Systems and Columns to the Official Standards

    No Cross-contamination

    • Customer results according to ISO 21675:2019, which cover 47 PFAS analytes 
    • Reporting limit: ≥ 0.2 ng/L
    • Cross-contamination with extremely highly contaminated samples (e.g. 4500 ng/L) only around 0.2%
    • Duration for 24 samples 6.5 hours, with an additional, customer-specific rinse ~ 8 hours
    EXCELLENT RESULTS

    Recovery Rates

    The results show that reliable and robust PFAS analysis can be performed via fully automated sample preparation on the FREESTYLE XANA PFAS robotic system in combination with EluCLEAN®PFAS SPE cartridges. The recoveries shown are very well within the accepted range of 70 - 130%. The precision is also well within the required range, with RSDs of 10% or less. Further results can be found in the following Application Note (AN0054):

    Recovery rates of 18 PFAS (listed in US EPA Method 537.1) + 4 surrogates from drinking water with EluCLEAN® SDVB columns (n = 6)